Determination of proper peaking time for Ultra-LEGe detector


KARABIDAK S. M., Kaya S., ÇEVİK U., Celik A.

RADIATION MEASUREMENTS, cilt.46, sa.4, ss.446-450, 2011 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 46 Sayı: 4
  • Basım Tarihi: 2011
  • Doi Numarası: 10.1016/j.radmeas.2011.01.023
  • Dergi Adı: RADIATION MEASUREMENTS
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.446-450
  • Anahtar Kelimeler: Pile-up, Peaking time, Resolving time, Dead time, Ultra-LEGe detector
  • Karadeniz Teknik Üniversitesi Adresli: Evet

Özet

Reducing count losses and pile-up pulse effects in quantitative and qualitative analysis is necessary for accuracy of analysis. Therefore, the proper peaking time for particular detector systems is important. The characteristic X-rays emitted from pure some elements were detected by using an Ultra-LEGe detector connecting a Tennelec TC 244 spectroscopy amplifier at different peaking time modes. Overall pulse widths were determined by an HM 203-7 oscilloscope connecting amplifier. The proper peaking time for Ultra-LEGe is determined as 3.84 mu s. (C) 2011 Elsevier Ltd. All rights reserved.