Determination of proper peaking time for Ultra-LEGe detector


KARABIDAK S. M. , Kaya S., ÇEVİK U. , Celik A.

RADIATION MEASUREMENTS, cilt.46, ss.446-450, 2011 (SCI İndekslerine Giren Dergi) identifier identifier

  • Cilt numarası: 46 Konu: 4
  • Basım Tarihi: 2011
  • Doi Numarası: 10.1016/j.radmeas.2011.01.023
  • Dergi Adı: RADIATION MEASUREMENTS
  • Sayfa Sayıları: ss.446-450

Özet

Reducing count losses and pile-up pulse effects in quantitative and qualitative analysis is necessary for accuracy of analysis. Therefore, the proper peaking time for particular detector systems is important. The characteristic X-rays emitted from pure some elements were detected by using an Ultra-LEGe detector connecting a Tennelec TC 244 spectroscopy amplifier at different peaking time modes. Overall pulse widths were determined by an HM 203-7 oscilloscope connecting amplifier. The proper peaking time for Ultra-LEGe is determined as 3.84 mu s. (C) 2011 Elsevier Ltd. All rights reserved.